HOME > ECG/EKG Monitor >Pocket ECG monitor
| Features
-
Portable design, compact in size with low weight, easy for carry.
-
Can be powered by an external DC power supply or a built-in rechargeable lithium battery.
-
Support synchronous acquisition and display of 6-lead / 12-lead waveform, as well as heart rate detection.
-
Provide ECG algorithm to automatically analyze the acquired ECG waveform, output measured values and diagnosis results.
-
Support auto mode and R-R mode and support automatic pacing detection and marking.
-
Provide 4 sampling modes: pre-sampling, real-time sampling, periodic sampling and trigger sampling.
-
Support ADS (Anti-drifting system) and EMG (electromyograph) interference.
-
Freeze the ECG waveform on the screen.
-
Auto-saving function: save the ECG data when the report is printed.
-
Support wireless transmission of ECG data via Wi-Fi network.
Light and Portable
About the size of a smartphone, Viatom Pocket ECG Machine is portable, light and easy to use.The pocket ECG machine only weighs 250g. Compare to traditional heavy ECG machine, our small ECG is a great choice for bedside monitoring.
Glasgow ECG Interpretation Algorithm
The Glasgow ECG Interpretation Algorithm, developed at the University of Glasgow, UK enables automated means of providing ECG analysis complementing the role of a physician. The algorithm accounts for age, gender and race and compares the ECG with historic data to arrive at specific interpretations.The Glasgow ECG Interpretation algorithm meets all the IEC 60601-2- 51 requirements and ISO 9001:2008 standards.
Why is Glasgow special?
-
Helps in QTc measurements facilitating assessment of cardiac risk.
-
Highly effective in interpreting STEMI (ST Segment Elevation Myocardial Infarction).
-
Can utilize V4R for neonates and children.
-
Offers short diagnostic reports for hospitals and detailed reports for primary care settings.
-
Critical values” included in diagnostic reporting template.
-
Pocket ECG machine combines complete data with sophisticated proprietary algorithms for comprehensive statistical analysis.
Reduct ECG Signal Noises by Filters
Pre-sampling, real-time sampling, periodic sampling, triggered sampling and automatic RR analysis modes to provide various measurements.
Drift filtering, EMG filtering, AC filtering and low-pass filtering makes the ECG signal stable and noise free, cause no distortion to the natural shape of the ECG waveform.
Application
Hospitals
Rural and remote areas
Corporates
Home
Primary health care
Diagnostic Labs
| Specification
Built-in rechargeable lithium-ion battery | 6000mAh for 24 hours recording |
Display | 4.46 inches, color LCD touch screen
Resolution: 480 × 854 pixels |
Weight | 250g, including the main unit and battery |
Dimensions | 134mm × 74mm × 17mm |
Minimum detectable signal | 20μVp-p |
ECG input signal range | ≤±5mVp-p |
Input impedance | ≥50MΩ (10Hz) |
Calibration voltage | 1mV±5% |
Depolarization voltage | ±500mV |
Noise | ≤30μVp-p |
Pacing pulse display | Pacing pulse with amplitude of 2mV~250mV, duration of 0.1ms~2.0ms, rise time of less than 100µs, and frequency of 100/min can be displayed on the ECG recording. |
Speed | 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s
Accuracy: ±3% |
Lowpass filter | 75Hz, 100Hz, 150Hz, Off |
ADS filter | 0.05Hz, 0.32Hz, 0.67Hz |
EMG filter | 25Hz, 35Hz, 45Hz, Off |
AC filter | 50Hz, 60Hz, Off |
Sensitivity | 2.5mm/mV, 5mm/mV, 10mm/mV, 20mm/mV
Accuracy: ±5% |
Frequency response | 0.05Hz~150Hz, 10Hz |
Time constant | ≥3.2s |
Common mode rejection ratio (CMRR) | ≥100dB (AC filter on) |
Sampling rate | 2000 samples / sec |
Leads | 6/12-lead synchronous acquisition and analysis |
A/D conversion | 24 bits |
HR Measurement range | 30bpm~300bpm |